    Product Name: HT-LH306 Humidity& Oxygen Transmitter
    器特点
    Instrument Features

    a. It is designed in split type, and the probe can be used in occasions whose temperature is within 200℃;

    b. The original imported zirconia sensor has a high measurement accuracy, strong anti-interference capacity, a long service life and stable performance, etc.;

    c. Polymeric membrane is built inside of the probe, which could prevent fluid media affecting or even damaging the sensor;

    d. The test cavity is made of 316L and is strongly corrosive-resistant;

    e. Automatic temperature compensation could eliminate the impact of environment temperature changes on the measurement accuracy;

    f. User could set upper and lower limits of humidity or oxygen of 2 ways of 4-20mA analog output;

    g. External display module can be attached to make the operation intuitive and convenient (optional);

    h. Wall mounting is easy and convenient to install.

    术参数
    Technical Parameters

    l Measurement Principle: Zirconia

    l Display Mode: External display mode can be equipped, 4 digit 16-segment alphanumeric LED display (optional)

    l Measurement Range: O2: 0~25.00%, H2O: 0~40.00%VOL

    l Measurement Accuracy: O2≤±2.5%FS

    H2O≤±2%FS

    l Repeatability: O2≤±1.5%FS

    H2O≤±2%FS

    l Resolution: 0.01%

    l Response Time: T90<30S

    l Output Port: 2 ways of 4-20mA (Non-isolated output, load resistance<1KΩ)

    2 ways of programmable alarm output (Dry contact type, maximum load 220VAC/2A)

    Display module port

    l Communication Mode: RS485 (default) or RS232

    l Power Supply: DC24V±10%, 1.5A

    l Environment Temperature: -10℃~+60℃

    l Storage Environment Humidity: <90%RH, non-condensing

    l Operation Environment Humidity: <100%RH, non-condensing

    l Sample Gas Temperature: 0~200℃ (tolerable temperature of probe)

    l Sampling Mode: Gas in

    l Gas Inlet Pressure: Relative pressure ±5kPa, stable pressure (control the pressure within the recommended limits to make sure the flow is within 400~600mL/min)

    l Gas Outlet Pressure: Free venting (under safe conditions)

    l Background Gas: Flue gas (except for strong-corrosive and reducing gas)

    l Sampling Port: NPT1/8-Φ6 cutting ferrule or fitting 316L (one out of two)

    l Size: 165mm×173mm×52mm (H×W×D)

    l Installation Size: 150mm×100mm (H×W)

    l Sensor Service Life: >36 months (under normal conditions of use)

    l Cable Length: 3 meters (standard)

    l Instrument Weight: Net weight 2.0Kg

    l Installation Method: Wall mounting

    用场合
    Application Scenarioses

    l The measurement of high-temperature humidity in the exhaust of oil-fried and coal-fried boiler (CEMS);

    l The measurement of humidity in food processing;

    l The measurement of humidity in timber, building material and paper making industry;

    l The measurement and analysis of water vapor content before and after desulfurization and denitration of flue gas in thermal power plant, chemical plant and steel plant.

    货须知
    Ordering Instructions

    l Measurement media

    l Display module (optional)

    l Corresponding measurement range of analog output

    Contact Number

    028-85044052/053